E-PSM(P)W Specifications
7-5
Document no. 2044677-001
Pulse rate from arterial pressure
Measurement range 30 to 250 bpm
Resolution 1 bpm
Accuracy ±5 bpm or ±5% whichever is greater
2.2.6 Respiration
The EMC immunity of the respiration measurement has been tested with 1 Vrms and 1 V/m.
This level has been used for optimizing the immunity of the respiration measurement to damp
the operating frequency of the electrosurgery equipment.
NOTE: Impedation respiration measurement is intended for patients over three years old.
Measurement range 4 to 120 breath/min
Accuracy ±5 breath/min or ±5%, whichever is greater
Resolution 1 breath/min
Gain range 0.1 to 5cm/ohm