Software download instruction
A-1
Document no. 2044677-001
APPENDIX A: Software download instruction
Overview
The ProgMon service download tool is a Windows-based software download application that
installs the main software and language files for B30 patient monitor.
WARNING The installation must only be done while the Monitor is not connected to a
patient in order to protect the patient and user from electrical shock.
CAUTION All user settings will be lost after installation of new monitor software. Save
user settings before starting software installation.
CAUTION Make sure that the monitor is connected to AC power source, or the batteries
are fully charged. Lost of power during software installation may result in
software installation failure.
System requirements
• PC with Microsoft Windows XP
• B30 software download CD
− B30 monitor main software
− B30 language files
− ProgMon software download tool
− B30 software download instruction
• CAT-5 Straight through network cable
• B30 software download license file
Software installation
NOTE: You can save your settings in Save Modes menu before software installation.
B30 monitor main software installation
1. Connect PC and B30 monitor into the same network with a network cable
2. Close the firewall and set the IP Address as 126.1.1.2, set the subnet mask as 255.0.0.0 in
PC
3. Turn on the B30 Monitor and enable software download function
Monitor Setup
Install/Service (16-4-34)
Service (26-23-8)
SW Settings
Set SW download to enable
The B30 monitor enters into software download mode.