Chapter 8: Specications
Performance Specications
Spectral Irradiance (bare beroptic pad)*
*Using an Ohmeda Medical BiliBlanket Meter II
Large Fiberoptic Pad – 49 µW•cm
-2
•nm
-1
(+/- 25%)
9-point check
Small Fiberoptic Pad – 70 µW•cm
-2
•nm
-1
(+/- 25%)
6-point check
NOTE: When the BiliSoft beroptic light pad is
inserted into a BiliSoft Pad Cover or BiliSoft Nest,
the nominal spectral irradiance is 35 µW•cm
-2
•nm
-1
(large pad) and 50 µW•cm
-2
•nm
-1
(small pad)
Wavelength 430-490 nm (peak 440-460 nm)
LED module estimated life** Under continuous use, tested at room temperature,
a typical LED module will run for approximately 8,000
to 10,000 hours before the light intensity drops 25%.
Sound level < 44 dB(A) at 1 meter
X-ray X-ray compatible
**The LED module life may vary when used in the actual clinical environment. Factors such as duty cycle and ambient temperature may
impact the life of the LED module. Measure the irradiance and replace the LED module when the BiliSoft system is below specications.
Physical Specications
Light box (W × H × L) 16.5 × 21 × 16.5 cm
Light box weight (excluding beroptic pad) < 2.5 kg
Fiberoptic Pad weight < 1.1 kg
Fiberoptic light pad, small 15 × 30 cm (light-emitting area)
Fiberoptic light pad, large 25 × 30 cm (light-emitting area)
Fiberoptic cable length 137 ± 5 cm