34 M1110150 006 © 2007 by General Electric Company. All rights reserved.
Chapter 9: Technical Reference
Large Fiberoptic Light Pad
Effective Surface Area
15cm.
Small Fiberoptic Light Pad
Effective Surface Area
30cm.
25cm.
30cm.
EMC Guidance
Safety Standards: IEC 60601-1, IEC 60601-2-50, EMC Standards IEC 60601-1-2
Table 1
Guidance and manufacturer’s declaration - electromagnetic immunity
The BiliSoft Phototherapy System is intended for use in the electromagnetic environment specied below.
The customer or the user of the BiliSoft Phototherapy System should assure that it is used in such an
environment.
Immunity Test IEC 60601 Test Level Compliance level
Electromagnetic
environment-guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 6kV contact +/- 8kV
air
+/- 6kV contact
+/- 8kV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material the
relative humidity should
be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
+/- 2kV for power supply
lines.
+/- 1kV for input/output
lines.
+/- 2kV for power supply
lines.
+/- 1kV for input/output
lines.
Mains power quality
should be that of a typical
commercial or hospital or
domestic environment.