1mAs~80mAs  @  200V,  220V,  230V,  240V;  current: 20mA, stationary;  the 
obtainable highest X-Ray tube voltage is 110kV; 
1mAs~40mAs  @  100V,  110V,  120V;  current:  10mA,  stationary;  the 
obtainable highest X-Ray tube voltage is 110kV. 
•  Film:  
Voltage: 40~110kV;  
mAs: 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13, 16, 20, 25, 32, 40, 50, 63, 80 mAs @ 200V, 220V, 
230V, 240V; 
mAs: 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13, 16, 20, 25, 32, 40 mAs @ 100V, 110V, 120V. 
•  Pulse Fluoro: 1/2/4/8pps 
•  The lowest Current Time Product or the combination of Loading Factors: 40kVp, 1mAs. 
•  Fluoroscopy Duty Cycle:  
80 kVp @ 1.0 mA continuous  
•  Film Mode Duty Cycle: 
110 kVp @ 20 mA, 4 seconds (80 mAs)  
Once per 5 min (1.3%) continuous  
Once per 2min (3.3%) for 10 exposures 
 
•  kVp Accuracy                         7% 
•  mA Accuracy                          10% 
•  mAs Accuracy                        (10% + 0.2mAs) 
•  Linearity Film mode linearity < 0.18 
•  Reproducibility Coefficient of Variation (COV)  < 0.045 
•  Measurement Basis for Technique Factors 
kVp - The peak value of high voltage generator output in the interval after a 20 mS delay period to 
the end of the exposure.  
mA - The time average of the current flow into the high voltage cable/X-ray tube assembly, 
beginning at the point where kVp crosses the 35 kVp level.  
Time - Measurement of exposure time begins when the kVp crosses the 35 kVp level ( 80% of kVp 
selected).  
mAs - The time integral of mA as defined above.