Technical Reference
Page 12-6
12.3. Specification of Key Components
12.3.1. X-Ray Generator
• Nominal electric power:
2.0kW (100kV, 20mA, 0.1s)
• Maximum electric power:
660W (110kV, 6mA; 55kV, 12mA)
1.76 kW (110kV, 16mA) @ 200V, 220V, 230V, 240V;
1.1 kW (110kV, 10mA) @ 100V, 110V, 120V
2.2 kW (110kV, 20mA) @ 200V, 220V, 230V, 240V;
1.1 kW (110kV, 10mA) @ 100V, 110V, 120V.
• High frequency generator: 40kHz
• Voltage: 40kV~110kV
When the X-Ray tube current reaches its highest value of 4mA, the
obtainable highest X-Ray tube voltage is 110kV.
When the X-Ray tube current reaches its highest value of 12mA, the
obtainable highest X-Ray tube voltage is 55kV.
When the X-Ray tube current reaches its highest value of 2mA, the
obtainable highest X-Ray tube voltage is 110kV.
When the X-Ray tube current reaches its highest value of 16mA, the
obtainable highest X-Ray tube voltage is 110kV @ 200V, 220V, 230V,
240V.
When the X-Ray tube current reaches its highest value of 10mA, the
obtainable highest X-Ray tube voltage is 110kV @ 100V, 110V, 120V.
20mA, stationary @ 200V, 220V, 230V, 240V, the obtainable highest X-Ray
tube voltage is 110kV;
10mA, stationary @ 100V, 110V, 120V, the obtainable highest X-Ray tube
voltage is 110kV.
0.1 mA~4 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 4mA.
0.2 mA~12 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 6mA.
0.1 mA~2 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 2mA.
0.2 mA~16 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 16mA @ 200V, 220V,
230V, 240V.
0.2 mA~10 mA; when the X-Ray tube voltage reaches its highest value of