4.1.  Overview 
This chapter describes: 
•  Start Up  
•  C-Arm Controls 
•  Workstation Controls 
•  Image Quality Control 
4.2.  Start Up  
4.2.1.  Power On 
Press the Power On/Off key on the Workstation to start up Workstation, use Workstation to view images. 
The C-Arm receives power through an interconnect cable attached to the Workstation, connect C-Arm to 
produce images.  
At  initial  setup  or  when  removing  from  storage,  stabilize  for  24  hours  at  ambient 
temperature and humidity before applying power. Failure to observe this warning will result 
in damage to equipment. 
Remove  any devices  from USB  connector  before booting
,
otherwise  system booting may 
fail. 
 
1.  Plug the Workstation power plug into a properly rated receptacle. Refer to the Chapter 12 “Technical 
Reference” for information about power requirements.  
 
If the system power off automatically due to power lost, user needs to power on the system 
after power restored to proceed. 
Brivo OEC 715/785/865 Mobile C-Arm X-Ray product provides a UPS to protect data loss. If 
the  power  lost  during  operation,  UPS  battery  will  supply  power  to  computer  and  right 
monitor until system shut down automatically.  
Do  not  disconnect, repair  or  replace  UPS  without permission.  Only  a  GE  qualified  service 
engineer can perform repair and replacement. The life cycle of the UPS is about 3-5 years in 
normal use. It may be different in different conditions.  
To use the system for the first time, it is recommended to boot the system for continuous 8  
hours  to  ensure  that  the  UPS  is  fully  charged,  otherwise  if  the  system  power  down 
accidentally, there is risk that the system will not safely power off; during the normal work, 
when  the  system  power  down  accidentally,  the  UPS  play  a  protective  role  to  ensure  the 
system safely power off, and it is recommended to boot the system 8 hours in advance to 
ensure that the UPS is fully charged the next time; an additional charge of the UPS is not 
necessary when the system does not abnormally power off.