Introduction and Safety
 
Page 1-22 
   
 
Guidance and manufacturer’s declaration - Electromagnetic Immunity 
The  Brivo  OEC  715/785/865  Mobile  C-arm  X-ray  Product  is  intended  to  use  in  the  electromagnetic  environment 
specified below. The purchaser or operator of the Brivo OEC 715/785/865 Mobile C-arm X-ray Product should assure 
that it is used in an electromagnetic environment as described below: 
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment 
IMMUNITY
 
TEST LEVEL 
(V/m) 
Pulse 
modulation 
b)
 
18 Hz 
FM 
c) 
5 kHz deviation 
1 kHz sine 
Pulse 
modulation 
b)
 
217 Hz 
GSM 800/900, 
TETRA 800, 
iDEN 820, 
CDMA 850, 
LTE Band 5 
 
Pulse 
modulation 
b)
 
18 Hz 
GSM 1800; 
CDMA 1900; 
GSM 1900; 
DECT; 
LTE Band 1,  3, 
4, 25; UMTS 
 
Pulse 
modulation 
b)
 
217 Hz 
Bluetooth, 
WLAN, 
802.11 b/g/n, 
RFID 2450, 
LTE Band  7 
 
Pulse 
modulation 
b)
 
217 Hz 
Pulse 
modulation 
b)
 
217 Hz 
NOTE   If necessary to  achieve the IMMUNITY TEST  LEVEL, the  distance  between the  transmitting antenna  and  the 
ME  EQUIPMENT  or ME  SYSTEM  may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. 
a)  
For some services,  only the uplink frequencies  are included. 
b)  
The carrier shall be modulated using a 50 % duty cycle square wave signal. 
c)  
As an alternative  to FM modulation, 50 % pulse  modulation at  18 Hz may be  used  because  while  it does  not  represent 
actual modulation, it would be worst case.