Usual Operations
Page 5-18
Country/
Region
MODE
(Dose measured at 30 cm above the Image Intensifier)
Fluoro
kVp: 40 to 110
mA: 0.1 to 4
Pulsed Fluoro
kVp: 40 to 110
mA: 0.1 to 4
High Level
Fluoro
kVp: 40 to 110
mA: 0.2 to 12
Pulsed High
Level Fluoro
kVP:40 to 110
mA: 0.2 to 12
Digital Spot
kVp: 40 to 110
mA: 0.2 to 16
Film
kVp:40 to 110
mA: 20
mAs:1 to 80
Algeria
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Argentina
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Australia
<50 mGy/min
<50mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Austria
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Bangladesh
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Belgium
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Bolivia
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Bosnia
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Brazil
<50 mGy/min
<50mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Brunei
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Bulgaria
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Cambodia
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Canada
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
Chile
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy
China
<88 mGy/min
<88 mGy/min
<100 mGy/min
<100 mGy/min
<2mGy
<20mGy

Table of Contents

Other manuals for GE Brivo OEC 865

Related product manuals