Introduction
2002938-002E Cardiolab® II Plus Amplifier Manual 1-13
WARNING
MAINTENANCE — Use only Catheter Input Module
(CIM) #301-00202-08 or #30100203-08 with the
CardioLab II Plus Amplifier.
Check the electrical integrity of the signal connections in
the catheter input modules and cabling yearly.
WARNING
POWER SUPPLY — Connect the system to a power
outlet equipped with emergency power.
The supply line must be fused and capable of delivering
20A.
Route the AC power cable away from moving parts.
Connect only to a properly earth grounded outlet.
WARNING
REPLACEMENT PARTS — Replace only with the same
type and rating of fuse.
WARNING
INDICATIONS OF USE — This device is intended for
use under the direct supervision of a licensed health care
practitioner.
CAUTION
RESTRICTED SALE — Federal USA law restricts this
device to sale by or on the order of a physician.
CAUTION
DATA ACCURACY — DO NOT display any channels
which are based on unused catheter input modules.
Erroneous values may occur if unused catheter inputs are
not properly connected.