EasyManuals Logo

GE Corometrics 172 Service Manual

GE Corometrics 172
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
Revision C 170 Series Monitor 6-25
2000947-004
Functional Checkout Procedure: Dual Heart Rate Test (Non-Pattern)
Table 6-6. Model 325 Simulator Settings for FECG/US Test
Section Switch Setting
FECG/MECG
Main RATE
Rate 120 BPM
Mode FECG
QRS Amplitude 50 µV
QRS Polarity +
ULTRASOUND/FMD
Mode US
Level MED
Rate RAMP
General Pattern Memory OFF

Table of Contents

Other manuals for GE Corometrics 172

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 172 and is the answer not in the manual?

GE Corometrics 172 Specifications

General IconGeneral
BrandGE
ModelCorometrics 172
CategoryMedical Equipment
LanguageEnglish

Related product manuals