GE CASE V6.7 - User Manual

GE CASE V6.7
342 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
GE Healthcare
CASE™ V6.7
Software Version 6.73
Operator’s Manual
2060290-201 ENG Revision C

Table of Contents

Questions and Answers

  • G
    George CaldwellAug 8, 2025
    What to do if my GE CASE V6.7 Medical Equipment says 'Not enough free memory on hard drive'?
    • D
      denisegillespieAug 8, 2025
      If your GE Medical Equipment displays a 'Not enough free memory on hard drive!' message, you need to increase the available storage. You can achieve this by compressing exercise test data or deleting unnecessary examinations. Also, ensure that both the hard drive containing your database and the one with your operating system have sufficient free space, with at least 100 MB available in the temporary directories.
  • M
    Matthew RowlandAug 12, 2025
    How to fix ECG grid not visible in PDF file from GE CASE V6.7?
    • A
      April KellyAug 12, 2025
      If the ECG grid isn't showing up in your PDF file from the GE Medical Equipment, configure Acrobat Distiller. For Acrobat V6.0: Go to Start -> Settings -> Printer -> Adobe PDF -> menu Printer -> Document Defaults -> menu Adobe PDF Settings -> disable 'do not send fonts to distiller'; menu Paper/Quality -> button Advanced -> Graphic and select the field 'Download True Type Font as Softfont'. For Acrobat V5.0: Follow the same steps. If the problem persists, change the grid format in System Configuration -> General tab -> Printer -> deselect 'Dotted line' -> select grid format. If the grid lines are too strong, use a shade of gray instead of black.
  • B
    blevinsduaneAug 17, 2025
    What to do if GE CASE V6.7 shows 'This document failed to print'?
    • B
      Brittany RosalesAug 17, 2025
      If you encounter a taskbar popup stating 'This document failed to print' with your GE Medical Equipment, and you can't delete the print job within the application, try deleting it directly from the print queue. Go to Start - Printers and Faxes - double-click Marquette Thermal - right-click the print job - left-click Cancel to delete the print job.
  • T
    Travis HuntAug 23, 2025
    Why GE CASE V6.7 patient record cannot be opened and is reported to be locked?
    • E
      erikchavezAug 23, 2025
      In a network environment, a patient record on your GE Medical Equipment may be locked if the client or server crashed.
  • A
    Aaron RayAug 28, 2025
    What to do if GE CASE V6.7 USB drive not found?
    • E
      elizabethwolfAug 28, 2025
      If your GE Medical Equipment isn't detecting the USB drive, try restarting the system.
  • M
    Michelle WilliamsNov 2, 2025
    Why won't my GE CASE V6.7 exercise test analysis start?
    • C
      Christopher GonzalezNov 2, 2025
      If the exercise test analysis on your GE Medical Equipment isn't starting, it could be due to several reasons. First, ensure that all the necessary electrodes are properly attached; if you still want to proceed, try clicking the 'Relearn medians' button. The device might be using unsuitable leads, so manually select appropriate ones. Check that the screen diagonal setting in the System Configuration is correct. For smaller screens (below 17 inches), make sure the Windows taskbar isn't set to always be on top. Verify that you're using the correct screen resolution with Small Fonts in Windows. Also, confirm that the ergometer model and port are correctly selected. Inspect the connection cable and connectors for any issues. Try turning the ergometer off and on again, and restart the GE Medica...

Summary

1 Introduction

Safety Information

Covers intended use, general information, definitions, classification, symbols, supplies, biocompatibility, maintenance, legal notice, security.

2 Equipment Overview

CASE System

Describes the front and rear views of the CASE system, including components like monitor, keypad, and ports.

Putting the CASE System Into Service

Details connecting system components, mains connection, power up, and functional tests.

Function Keypad

Explains the functions of the keys on the keypad for operating the ergometer/treadmill and controlling tests.

3 Preparing the Patient

Applying Electrodes - The Basics

Explains the importance of electrode application for noise-free ECG and recommends GE Healthcare products.

4 Selecting a Patient Record and Test Type

Selecting a Patient Record

Guides on how to find and select a patient record from the local database to start a test.

Entering a New Patient

Provides detailed steps for entering new patient information into the system.

5 Resting ECG

Calculating Patient Risk

Explains how the system calculates risk for stroke and CHD based on patient data.

Recording a Resting ECG

Overview of the screen for verifying ECG signal quality and acquiring resting ECGs.

10-Second ECG with Analysis

Details how the system continuously saves and analyzes 10-second ECG segments.

Generating the Interpretation

Guides on how to open and edit the interpretation window using standard texts and acronyms.

6 Exercise Test

Before the Test

Overview of the Pre-Acquisition Screen, including lead check, tabs, and patient education.

During the Test

Overview of the Acquisition Screen and its components during the exercise test.

Pretest Phase

Details the stages of the pretest phase and operating steps for recording baseline ECGs.

Exercise Phase

Describes how the selected protocol controls the ergometer/treadmill during the exercise phase.

ST;Arrhy Window

Describes the ST/Arrhy window displaying median complex and current arrhythmias.

Recovery Phase

Describes advancing to the recovery phase by pressing the recovery button.

Test End Phase

Overview of the test end phase, including confirming test end and entering interpretation.

After the Test - Post Test Review

Overview of the Post Test Review screen, covering Test Summary, patient, test, measurement, and interpretation areas.

Full Disclosure ECG

Explains how to view the full disclosure ECG, including events and display options.

7 Ambulatory Blood Pressure Measurement

Setting Up the Ambulatory Blood Pressure Monitor

Overview of the Acquisition Screen for setting up ambulatory blood pressure measurements.

Downloading Data

Instructions for downloading data from the BP monitor to the PC.

Test Summary

Overview of the Test Summary screen for BP measurements, including patient, test, and measurement results.

8 Spirometry Test

Entering Test Information

Details the Test Information window, including Patient Information, Test/Personnel, and Medications tabs.

Conducting Spirometry Tests

Explains spirometry test procedures with SpiroSoft and LF 501 sensors, including measurement modes.

Flow-Volume Measurement - Operating Steps

Details the steps for performing flow-volume measurements, including patient explanation and maneuver.

Bronchodilator Test

Explains the procedure for bronchodilator tests, involving measurements before and after medication.

FVC Measurement

Instructions for performing Forced Vital Capacity (FVC) measurements.

Post Test Review

Overview of the Test Summary screen for spirometry, including patient, test, and measurement results.

Generating or Editing the Interpretation

Guides on how to open and edit the interpretation window using standard texts.

9 File Management

Selecting Patient Records

Guides on how to select patient records, filter them by various criteria, and perform actions like printing or generating reports.

Archiving, Exporting, Deleting Patient Records

Covers selecting patient records and performing actions like saving to archive, transferring to MUSE, or exporting.

11 DICOM Interface

Performing Procedures with the Order List (Scheduled Procedures)

Guides on performing procedures based on an order list, including patient selection and starting procedures.

12 System Settings

Resting ECG Setup

Covers settings for the resting ECG modality, accessible via the acquisition screen or System Configuration.

Exercise Test Setup

Covers settings for exercise tests, accessed via the pre-acquisition screen or System Configuration.

Ambulatory Blood Pressure Measurement Setup

Covers settings for ambulatory BP measurements, accessed via acquisition screen or System Configuration.

Spirometry Setup

Covers settings for spirometry, accessed via the acquisition screen or System Configuration.

Custom Setups, Factory Setup

Allows saving, loading, and deleting personal settings, and restoring factory defaults.

System Configuration

Overview of the System Configuration menu, accessed from the initial screen.

A Cleaning and Maintenance

Cleaning, Disinfection, and Maintenance

Guidelines for cleaning and disinfecting the equipment surface, cables, and monitor.

Maintenance

Covers routine checks and technical inspections required for system safety and performance.

B Miscellaneous

Exercise Test Interpretation (XTI)

Explains the XTI program's comparison of exercise measurements against thresholds for predicting morbidity/mortality.

Duke Treadmill Score

Details the Duke Treadmill Score computation, risk categories, and prognosis percentages.

Spirometry Reference Value Equations, Interpretation Modes, Measurements

Details reference value equations, interpretation modes, and definitions of spirometry test values.

Troubleshooting

Provides solutions for common problems encountered with the system, such as data transfer issues or incomplete displays.

D Technical Description

Technical Specifications - CASE System

Details performance specifications for signal processing, data acquisition, and display.

Table 1. Performance Specifications - Signal Processing

Details specifications for ECG analysis frequency, ST measurements, signal processing, and interpretation.

Table 2. Performance Specifications - Data Acquisition

Details specifications for data acquisition technology, sampling rate, and skew.

Table 7. Safety

Outlines certification, type of protection against electrical shock, patient leakage current, and maintenance frequency.

GE CASE V6.7 Specifications

General IconGeneral
BrandGE
ModelCASE V6.7
CategoryMedical Equipment
LanguageEnglish