Revision C CASE 7-7
2060290-201
Post Test Review
Post Test Review
Downloading Data
1. Turn the monitor off and on again.
2. Wait for the time to be displayed, then connect the monitor to the PC.
3. Select the patient.
Note
If the patient ID has been stored in TONOPORT V devices of
version 2.0 and later, you are not required to select the patient.
Simply click Select and you will see the acquisition screen.
Be sure to download the data to the same database from which
you read the patient ID.
4. Click Connect.
5. Click Download Data. After the download, the Test Summary tab
appears.
Note
Do not forget to delete the data stored in the monitor. This is to
ensure that the monitor does not contain data from a previous
test when used on a new patient.
When the message Inconsistent data in ambulatory BP device!
Procedure cannot be saved! appears, turn the monitor briefly off
and on again.