Revision C CASE 12-3
2060290-201
Resting ECG Setup
Resting ECG Setup
The settings specific to the resting ECG modality can be entered either
from the resting ECG acquisition screen where you click Setup to display
the setup menu, or from the General tab of the System Configuration (see
“General Tab” on page 12-40).
Acquisition Tab
The Acquisition tab with the acquisition settings is already on top.
A Click to print the resting ECG setup.
B Click to display the other setup tabs.
C
Select or deselect display of the 1-mV calibration pulse.
D Select the speed and gain (screen).
E Select the screen format.
F
Select to hear the lead off alarm. If selected, the system will emit a beep
when an electrode becomes disconnected.
G
Select to enable the Cubic Spline algorithm (automatic compensation of
baseline fluctuation). Cubic Spline causes a signal delay of approx. 2.
seconds.
H
Click to enable the thermal writer.
I Select the thermal writer speed and gain.
J Select the number of rhythm leads to be recorded with the thermal writer.
K Select filter frequency (ECG).
L Click to edit the Reason for Test library (see next page).
M Click to edit the Interpretation library.
N Click to edit the Medications library.