12-8 CASE Revision C
2060290-201
Resting ECG Setup
Miscellaneous Tab
A
Select the check boxes to enable the following functions:
Preview before Analysis: Full-screen display of the 10-second ECG, analysis only
after confirmation.
Print after Analysis: Automatic printout after acquisition of the 10-second ECG.
Post Test Review after Analysis: Automatic display of the test report after
acquisition and analysis of the 10-second ECG.
Start Full Disclosure after Analysis: After acquisition of the 10-second ECG the
system starts saving the full disclosure data.
Pace Enhancement: All pacer pulses will be displayed with an amplitude of 0.5
mV.
QRS Beep: The system emits a beep each time it detects a QRS complex (only
with CAM-14 acquisition module).
Axis Graph: An axis graph is shown in Tabular Summary and in the Median
Report.
B Select sampling rate for full disclosure ECG.
C Enter prompt for Extra Question 1 (appears in Test Information, Test/Personnel).
D Select answer format for prompt 1.
E Select sagittal plane (right/left) for vector loop display.
F 12SL analysis program:
Select check box to display the corresponding statements in the Test Summary
(ACI-TIPI (Acute Cardiac Ischemia - Time Insensitive Predictive Instrument) is a
mathematically-based decision aid which has been shown to be useful in critical
care situations where speed of diagnosis can be crucial).
G HEART analysis program.