EasyManuals Logo

GE CASE V6.7 User Manual

GE CASE V6.7
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #240 background imageLoading...
Page #240 background image
12-8 CASE Revision C
2060290-201
Resting ECG Setup
Miscellaneous Tab
95A-R
A B C D F GE
A
Select the check boxes to enable the following functions:
Preview before Analysis: Full-screen display of the 10-second ECG, analysis only
after confirmation.
Print after Analysis: Automatic printout after acquisition of the 10-second ECG.
Post Test Review after Analysis: Automatic display of the test report after
acquisition and analysis of the 10-second ECG.
Start Full Disclosure after Analysis: After acquisition of the 10-second ECG the
system starts saving the full disclosure data.
Pace Enhancement: All pacer pulses will be displayed with an amplitude of 0.5
mV.
QRS Beep: The system emits a beep each time it detects a QRS complex (only
with CAM-14 acquisition module).
Axis Graph: An axis graph is shown in Tabular Summary and in the Median
Report.
B Select sampling rate for full disclosure ECG.
C Enter prompt for Extra Question 1 (appears in Test Information, Test/Personnel).
D Select answer format for prompt 1.
E Select sagittal plane (right/left) for vector loop display.
F 12SL analysis program:
Select check box to display the corresponding statements in the Test Summary
(ACI-TIPI (Acute Cardiac Ischemia - Time Insensitive Predictive Instrument) is a
mathematically-based decision aid which has been shown to be useful in critical
care situations where speed of diagnosis can be crucial).
G HEART analysis program.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CASE V6.7 and is the answer not in the manual?

GE CASE V6.7 Specifications

General IconGeneral
BrandGE
ModelCASE V6.7
CategoryMedical Equipment
LanguageEnglish

Related product manuals