B-22 CASE Revision C
2060290-201
Miscellaneous: Remote Service
Remote Service
In conjunction with the InSiteExC application, CASE is a system with
remote service capability. For InSiteExC installation instructions, please
refer to the CASE Field Service Manual. You must have access to the
internet to be able to use the remote service capability.
Warning
Patient Hazard — Disconnect the patient from CASE
during remote servicing.
Virus Protection
Any anti-virus software installed on the CASE must be compatible with
the CASE operating system.
Automatic online updates should be disabled. Instead, we recommend
periodically updating the virus definition database via modem or TCP/IP
network, using the manual Live Update feature (once a week, if
possible).
Warning
Patient Hazard — Disconnect the patient from the
system when the modem is connected.