Revision C CASE 6-17
2060290-201
During the Test
Manual Control
Note
Cubic Spline/FRF algorithms can delay the rhythm report by
approximately 2 seconds when enabled.
Note
In the pretest phase, Shift + pretest or Shift + F1 allow you to return
to the previous stage.
Note
If a bicycle ergometer test is initiated with the exercise button, the
clock will start even when the patient has not yet begun to pedal.
Note
When the patient reaches the target HR, the HR reading turns red
and an audio signal sounds (if configured, see section “Miscellaneous
1 Tab” on page 12-24).
Note
Using the treadmill speed and grade buttons during a protocol puts
the treadmill in manual control. The system is no longer controlled
automatically by the protocol. The system will enter a new stage and
the phase name appearing in the Tabular Summary will be Manual.
comment
Press to enter comments to appear in the Tabular Summary.
enter BP
Press to enter BP readings.
12 lead
Press to start an ECG recording (12 leads, 5 seconds from memory, 5
seconds of real-time data).
recall Press to print a 10-second delayed recall report on the thermal writer.
rhythm
Press to start a continuous rhythm report on the thermal writer.
stop writer
Press to stop the writer or printer.
medians Press to print a medians report.
arrhy doc
Press to enable or disable the automatic documentation of
arrhythmias.
12SL analysis
Press to acquire a 12SL resting ECG with measurements and
interpretation (Pretest and Recovery phases only).
grade +
grade -
Press to change the elevation of the treadmill belt.
speed w+
speed w–
Press to change the belt speed or ergometer load.
exercise
Press to advance to the exercise phase.
hold stage
Press to hold the exercise test in the current stage.
start tmill/
STOP tmill
Press to start and stop the treadmill belt.