Revision C CASE A-3
2060290-201
Cleaning and Maintenance: Cleaning, Disinfection, and Maintenance
Cleaning, Disinfection, and Maintenance
Equipment Surface
Warning
Shock Hazard — Disconnect the power cord from the wall
outlet before cleaning or disinfecting the system.
Use a moist cloth to wipe the surface clean; do not allow liquids to
enter the system. All cleaning agents and disinfectants commonly
used in hospitals and containing up to 70 % alcohol are suitable. If
liquids have entered the system, notify service to have the system
inspected for damage before it is used again.
DO NOT use disinfectants with a phenol base or peroxide compounds
to disinfect the external surfaces.
Cables
Disconnect the cable from the system before cleaning or disinfection.
To disconnect cables, always pull on the plug, not on the cable.
Rub the cables clean with soap water, use a disinfectant for
disinfection. All cleaning agents and disinfectants commonly used in
hospitals and containing up to 70 % alcohol are suitable. Do not
immerse the cables in liquids.
Monitor
For cleaning and disinfection of the monitor housing, refer to the
information given above under „Equipment Surface“.
Clean the LCD monitor surface with a dry, lint-free, soft cloth. If
required, moisten the cloth with water or with the combination
cleaners for housing and screen Alklar, Sidolin manufactured by
Henkel.
Manufacturer:
Henkel KG a.A.
http://www.henkel.com