Revision C CASE 8-13
2060290-201
Conducting Spirometry Tests
FVC Measurement
1. On the acquisition screen click FVC Measure..
2. Explain the test maneuver to the patient.
3. Check the nose clip.
4. Click Start Test and ask your patient to perform the following
maneuver when the orange indicator on the sensor lights up:
Inhale completely through the sensor. Exhale maximally and as
quickly as possible, then inhale again completely and as quickly
as possible. Then your patient may resume normal breathing.