7-34 170 Series Monitor Revision C
2000947-004
Serviceable Assemblies: Testing a Repaired Transducer (TOCO or US)
Testing a Repaired Transducer (TOCO or US)
CAUTIONS
Following completion of the cable replacement procedure:
LEAKAGE TEST—Perform a leakage and dielectric test on the
transducer per applicable standards.
FUNCTIONAL TEST—Perform a functional response test on the
transducer.