9-6 170 Series Monitor Revision C
2000947-004
Maintenance: Preventative Maintenance Inspection
Cleaning
Clean the monitors exterior. Refer to “Cleaning the Monitor Exterior” on
page 9-2.
Clean the recorder printhead. Refer to “Cleaning the Printhead” on page 7-
46.
Clean the monitors accessories. Refer to “Cleaning the Tocotransducer,
Ultrasound Transducer, and Legplate” on page 9-3.
Clean the monitors interior using a hand-held vacuum.
Calibration
Adjust the printhead. Refer to “Aligning the Printhead” on page 7-43.
Calibrate the tocotransducer(s). Refer to “Tocotransducer Calibration” on
page 7-20.