User Manual Druck DPI 620G-IS
GE Digital Solutions
Druck DPI620 Genii
Intrinsically Safe
Calibrator
and Communicator Series
User Manual 116M5464
© 2016 General Electric Company. All Rights Reserved. Specifications are subject to
change without notice. GE is a registered trademark of General Electric Company. Other
company or product names mentioned in this document may be trademarks or
registered trademarks of their respective companies, which are not affiliated with GE
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the GE DPI620G-IS and is the answer not in the manual?

Overview

The Druck DPI620G-IS is a battery-powered, intrinsically safe calibrator and communicator designed for use in hazardous areas. It offers a comprehensive suite of electrical measure and source operations, alongside advanced digital communication capabilities for HART®, Foundation Fieldbus, and Profibus protocols. The device features a touch-screen display that can simultaneously show up to five different parameters, enhancing user efficiency and data visualization.

Function Description

The DPI620G-IS functions primarily as a versatile calibrator, capable of independent operations across five channels. Its core capabilities include:

  • Electrical Measure & Source Functions:
    • Current (measure & source)
    • Voltage (measure & source)
    • Thermocouple (measure & simulate)
    • Frequency (measure & source)
    • Resistance (measure & simulate)
    • RTD (measure & simulate)
    • Pulses (measure & source)
    • Observed (source): Allows manual entry of a value, such as a reading from a connected voltmeter.
  • Digital Communication:
    • HART® Communicator: Supports Universal and Common Practice commands (HART® revision 5 to 7) and devices with Device Descriptions (DD). It can operate as a Primary Master, controlling all communications, or as a Secondary Master, communicating with field devices within an existing HART® network.
    • Foundation Fieldbus Communicator: Implements the H1 protocol via an integrated H1 modem (available in DPI620G-IS-FF or DPI620G-IS-FFPB models). It supports connections to existing networks, stand-alone devices, and various network configurations.
    • Profibus Communicator: Utilizes an integrated modem for PROFIBUS PA Fieldbus implementation (available in DPI620G-IS-PB or DPI620G-IS-FFPB models).
  • Documentation Capabilities:
    • Data Logging: Records instrument readings for review and analysis, with options for periodic or key-press logging. Data is stored in a Comma Separated Variable (CSV) format, facilitating import into spreadsheet applications.
    • Analysis Function: Calibrates the transfer characteristic of a device under test by taking readings from two or more DPI620G-IS channels. One channel serves as the Reference, providing the input signal, while other channels measure the output. It calculates deviation in %Span or %Rdg and indicates pass/fail results based on tolerance limits.
    • Run Procedure: Executes calibration procedures downloaded from 4Sight or Intecal v10 software, including test points and ramp times.

Important Technical Specifications

  • Intrinsic Safety Classification: II 1 G Ex ib IIC T4 Gb (-10°C ≤ Ta ≤ +50°C).
  • Display: LCD color display with touch-screen.
  • Operating Temperature: -10 to 50°C (14 to 122°F).
  • Storage Temperature: -20 to 70°C (-4 to 158°F).
  • Ingress Protection: IP45 (for DPI620G-IS calibrator only). Note: This rating is for reliability, not safety, and the device is not suitable for prolonged UV exposure or permanent outdoor installation.
  • Humidity: 0 to 90% relative humidity (RH) non-condensing.
  • Shock/Vibration: MIL-PRF-28800F for class 2 equipment.
  • Pollution Degree: 2.
  • EMC: Electromagnetic compatibility: BS EN 61326-1:2006.
  • Electrical Safety: BS EN EN 60950-1:2010 + A2:2013 (Charger PSU Only).
  • Pressure Safety: Pressure Equipment Directive - Class: Sound Engineering Practice (SEP).
  • Battery: Lithium-Polymer battery (GE Part number: IO620G-IS-BATTERY).
    • Capacity: 5040 mAh (minimum), 5280 mAh (typical).
    • Nominal voltage: 3.7 V.
    • Charge temperature: 0 to 40°C (32 to 104°F). Charging stops outside this range.
    • Discharge temperature: -10 to 50°C (14 to 122°F).
    • Charge/discharge cycles: > 500 > 70% capacity.
  • Power Supply (IO620G-PSU): Input range 100 – 240Vac, 50 to 60Hz, 250mA, installation category CAT II. Operating temperature range 0°C to +40°C, storage temperature range -40°C to +70°C.
  • USB Client Connection: Must be connected to apparatus within a Safe Area with Um = 254V.
  • Channel 2, 15V, Uo loop supply: Remote sensors powered by this supply must be disconnected from all other power sources. Can connect to a FISCO system if power is from DPI620G-IS and electrical parameters are compatible.
  • Internal Loop Power (Fieldbus): Current limited to approximately 30mA.
  • Current Limits: Loop power current limit is 40mA.
  • Voltage Limits: Do not connect more than 30V CAT I between terminals or between terminals and ground.

Usage Features

  • Dashboard Navigation: Intuitive navigation by swiping the screen from top to bottom for main functions and from right to left for function-specific screens.
  • Custom Task Creation: Allows users to configure Channels 1 & 2, Pressure Channels (P1, P2), External (IDOS), and Communication (HART, Foundation Fieldbus, Profibus) settings.
  • Favourites: Users can save and copy frequently used tasks for quick access.
  • Function Utility Options: Includes Max/Min for measuring functions (displays minimum, maximum, and average values), Switch Test (detects switch opening/closing and displays hysteresis), and Relief Valve (tests cut-out response at defined thresholds, displaying max/min values).
  • Measurement Display Options: Offers "reduced view" for multiple channels and "expanded view" for detailed display of a selected channel.
  • HART® SDC Application: Provides a dedicated interface for HART® communication, including device polling, configuration viewing, editing values, and executing methods like self-test and analog trim.
  • HART® Offline Feature: Allows viewing, amending, creating, uploading, importing, and exporting HART® configuration files offline.
  • FOUNDATION™ Fieldbus and PROFIBUS PA Applications: Dedicated interfaces for scanning devices, viewing device profiles, navigating block navigation trees, displaying folder variables, and executing methods for configuration and calibration.
  • Function Finder: A search tool for Fieldbus and Profibus variables and device functions, simplifying navigation in complex devices.
  • On-screen Keyboard: Used for entering tags and other alphanumeric data.
  • Stylus: Included for precise touch-screen interaction.

Maintenance Features

  • Software and Firmware Upgrades: The device supports upgrades for the Operating System (OS), Application, HART, and CH1 FPGA software via downloadable .zip files from the manufacturer's website. The HART Device Library can also be updated.
  • Cleaning: The case and display should be cleaned with a lint-free cloth dampened with a weak detergent solution. Solvents or abrasive materials must not be used.
  • Instrument Return: For unserviceable instruments requiring repair, users must contact a GE Service Centre or Approved Service Agent to obtain a Returned Material Authorization (RMA) number. Information regarding product, serial number, defect details, and operating conditions is required.
  • Battery Replacement: The re-chargeable battery pack can be removed or replaced in a Hazardous Area. Recharging must be done in a Safe Area using only the GE specified charger and power supply.
  • Disposal: The product and its battery should not be disposed of as household waste. An approved organization for collection and recycling should be used, adhering to local health and safety procedures.
  • Calibration Procedures: GE provides calibration services traceable to international standards. If using an alternative facility, ensure it uses the specified standards. Calibration should be performed in a stable temperature environment (21 ± 1°C / 70 ± 2°F) after a minimum two-hour stabilization period. Specific calibration procedures are outlined for current, voltage, millivolts, TC mV, frequency, pulse amplitude, resistance, cold junction, and pressure indicator modules.