8 - [EN]  English K346 Issue 2
Operation - mA measurements
DPI 822 only. To measure a current:
1. Connect the instrument (Figure 5) and, if necessary, 
adjust the Set Up (Table 1).
2. Select the applicable mA input task from Select Task 
(Table 2/3) and, if necessary, adjust the Settings 
(Table 4).
Figure 5: Example configuration - mA measurement
Operation - Switch test
DPI 822 only. To do tests on a thermocouple switch:
1. Connect the instrument (Figure 6) and, if necessary, 
adjust the Set Up (Table 1).
2. Select the applicable switch test from Select Task 
(Table 2/3) and, if necessary, adjust the Settings 
(Table 5). The display shows the switch condition (open 
or closed) in the top right-hand corner.
3. Supply the output values to the system (Table 7).
• Example - “Nudge” output.
a. Use Edit (■
 ■) to set a value less than the switch 
value.
b. Use the ▲ ▼ buttons to change the value in small 
increments.
• Example - “Ramp” output.
a. Set “High” and “Low” values that are applicable to 
the switch value (Table 6). Then, to get an accurate 
switch value, set a long “Travel” period. 
b. Use Start/Stop (■
 ■) to start and stop the “Ramp” 
cycle.
4. If necessary, supply the output values in the opposite 
direction until the switch changes condition again.
The display shows the applicable values to open and 
close the switch.
5. To do the test again, press ESC to reset the values.
Figure 6: Example configuration - Switch test
Settings Settings
a) with external loop 
power
b) with internal loop 
power
Maximum: 30V
Start SettingsEdit Settings
a) Temperature b) mV