12 - [EN] English K346 Issue 2
Specification - Temperature ranges
Specification - mV range Specification - Electrical connectors (A2)
Thermocouple type Standard Range °F Range °C Accuracy °F * Accuracy °C *
K IEC 584 -454 .... -328 -270 .... -200 3.6 2.0
K IEC 584 -328 ... 2498 -200 ... 1370 1.1 0.6
J IEC 584 -346 ... 2192 -210 ... 1200 0.9 0.5
T IEC 584 -454 .... -292 -270 .... -180 2.5 1.4
T IEC 584 -292 ...... -94 -180 ...... -70 0.9 0.5
T IEC 584 -94 ..... 752 -70 ..... 400 0.6 0.3
B IEC 584 122 ..... 932 50 ..... 500 7.2 4.0
B IEC 584 932 ... 2192 500 ... 1200 3.6 2.0
B IEC 584 2192 ... 3308 1200 ... 1820 1.8 1.0
R IEC 584 -58 ....... 32 -50 ......... 0 5.4 3.0
R IEC 584 32 ..... 572 0 ..... 300 3.6 2.0
R IEC 584 572 ... 3216 300 ... 1769 1.8 1.0
S IEC 584 -58 ....... 32 -50 ......... 0 4.5 2.5
S IEC 584 32 ..... 212 0 ..... 100 3.4 1.9
S IEC 584 212 ... 3216 100 ... 1769 2.5 1.4
E IEC 584 -454 .... -238 -270 .... -150 1.6 0.9
E IEC 584 -238 ... 1832 -150 ... 1000 0.7 0.4
N IEC 584 -454 ........ -4 -270 ...... -20 1.8 1.0
N IEC 584 -4 ... 2372 -20 ... 1300 1.1 0.6
L DIN 43710 -328 ... 1652 -200 ..... 900 0.6 0.3
U DIN 43710 -328 ..... 212 -200 ..... 100 0.9 0.5
U DIN 43710 212 ... 1112 100 ..... 600 0.6 0.3
C 32 ... 2732 0 ... 1500 1.8 1.0
C 2732 ... 3632 1500 ... 2000 2.5 1.4
C 3632 ... 4208 2000 ... 2320 3.4 1.9
D 32 ... 3092 0 ... 1700 1.8 1.0
D 3092 ... 3992 1700 ... 2200 2.9 1.6
D 3992 ... 4523 2200 ... 2495 6.5 3.6
*Cold Junction (CJ) error (Maximum):
Range 50° ... 86°F (10 ... 30°C) = 0.4°F (0.2°C)
Add 0.01° CJ error / ° ambient temperature change for ranges:
14 … 50°F, 86 … 122°F
(-10 … 10°C, 30 … 50°C)
Range (mV) Impedance
-10 ... 100 < 0.2
Range (Measure) 0 to ±55 mA
Accuracy 0.02% of reading + 3 counts
Temperature coefficient
14 … 50°F, 86 … 122°F
(-10 … 10°C, 30 … 50°C)
0.0011% FS / °F
(0.002% FS / °C)
Switch detection Open and closed. 2 mA current.
Loop power output 24 V ± 10%
HART
®
resistor
250 (menu selection)
Connectors (A2) Three 0.16 in (4 mm) sockets