[EN] English - 3K346 Issue 2
To start - Power on or off
To turn the instrument on or off, press (A1 - item [1]). The
instrument does a self test and then shows the applicable
data.
When the power is off, the last set of configuration options
stays in memory. Refer to “Maintenance”.
To start - Set up the basic operation
Use the Set Up menu to set up the basic operation of the
instrument.
If there is additional data for a menu option, select
Settings (■■) to see the values that are set up. If
necessary, adjust the values.
Table 1: (Part of table) Menu options - Set Up
Table 1: (Part of table) Menu options - Set Up
To start - Select a task (Measure and/or simulate)
When the instrument is set up (Table 1), use the Select Task
menu to select the applicable task.
In Table 2/3, IDOS is a Universal Measurement Module
(UMM). If you attach a UMM to the communications port
(A1 - item [9]), the Select Task menu shows the applicable
IDOS options.
Table 2: Menu options - Select Task
Table 3 shows all the one and two function operations that
are available. If you attach a UMM, you can only use the
options that include IDOS.
1
Menu:
Select Task
23
Menu:
Set Up
45
(Table 2) (Table 1) []/[ ]
Options
(If applicable)
Description
DPI 822 only. To add a series resistor into the mA
circuit. You can then use this instrument together
with a HART
®
communicator to set up and
calibrate HART
®
devices.
... Scale To select the applicable international temperature
scale: IPTS 68 or ITS 90.
CJ ... To select the type of cold junction (CJ)
compensation.
Automatic: The instrument monitors the CJ
temperature and applies the necessary CJ
compensation.
Manual: Measure the CJ temperature and set the
applicable value. The instrument uses this value to
apply the necessary CJ compensation.
Additional data (Manual): Select Settings ( )
To select and set up the backlight facility + timer.
Additional data: Select Settings ( )
To select and set up the power off facility + timer.
Additional data: Select Settings ( )
To show the battery level (%).
To set the display contrast (%).
Increases %, decreases %
Options
(If applicable)
Description
To set the time + date. The calibration facility uses
the date to give service and calibration messages.
To set the language option.
To calibrate the instrument.
Additional data: Refer to “Calibration”.
To select and show the applicable status data.
(Software Build, Calibration Due date, Serial
Number, IDOS Information).
1
Menu:
Select Task
23
Display:
T1 output
(Table 2/3)
Sk1 = Edit
Sk2 = Settings
Options
(If applicable)
Description
T1 or
mV
An input measurement task:
T1 - Measure thermocouple temperature OR
mV - Measure thermocouple mV.
T1 or
mV
An output task:
T1 - Simulate thermocouple temperature OR
mV - Simulate thermocouple mV.
mA DPI 822 only. A mA measurement task.
mA(24V) DPI 822 only. A mA measurement task + the loop
power supply is on.
DPI 822 only. A switch test.
IDOS UMM only. An IDOS measurement task.
UPM only. A leak test.
To set up the way the instrument works.
Additional data: Refer to: Set Up (Table 1).