- 104 - eBike, eBike L, eBike EL                                     2018112-003 Rev J
Appendix C: Technical Specifications
safety standards: BF to IEC 60601-1
protection class: I to IEC 60-601
MDD classification: class IIa to 93/42 EEC
Environment:
operation: temperature: +10...+40 °C/50...104 °F
relative humidity 30 ... 75%, 90% no condensation
atmospheric pressure: 700 ... 1060 hPa
transport and storage: temperature: -40 ... +70 °C (-40 ... +158 °F)
relative humidity: 10 ... 95%, no condensation
atmospheric pressure: 500 ... 1060 hPa
Optional: BP Module
measuring principle: - indirect measurement to Riva-Rocci-Korotkov
- oscillometric
the resting BP readings of both methods are compared for plausibility
cuff connection: tubing with connector and jack plug (microphone)
measuring range:
systolic pressure: 40 to 300 mmHg
diastolic pressure: 40 to 300 mmHg
pulse: 35 to 230 P/min
accuracy of the (static)
pressure readout: +/-3 mmHg
resolution of the readout: +/- 1 mmHg
initial inflation pressure during the inflation phase, the inflation pressure automatically adapts to patient’s
BP
pump rate: between approx. 6 seconds (to 140 mmHg) and approx. 18 seconds (to 300
mmHg)
max. cuff pressure: 300 mmHg
cuff deflation method: pulse-dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
calibration: calibration with external pressure meter
artefact rejection: automatic artifact rejection and comparison of the resting BP readings of both
methods for plausibility