- 6 - eBike, eBike L, eBike EL 2018112-003 Rev J
4. Both couch adjustment motors
- Set both motors to both stop positions.
- Change couch position from horizontal (0°) to semi-recumbent (45°) and then back to the
horizontal position (0°).
- Check adjustment range for lateral tilt from the horizontal position (0°).
Expected result: Couch position changes to the expected positions Pass/Fail
5. Saddle adjustment motor
- Check all possible positions by moving the saddle to the top and to the bottom position.
- Return saddle to the middle position.
Expected result: Saddle position changes to the expected positions Pass/Fail
6. Speed and Noise
- Check speed indication on LCD and additional speed indication for test subject.
- Check noise level of idling drive unit (e.g., for grinding noises or noise from the bearing).
Expected result: Speed displayed on both displays and no noticeable noise Pass/Fail
7. NIBP module (if applicable)
- Check blood pressure measurement on test subject at rest.
- Check blood pressure readings on LCD.
Expected result: BP values are displayed and in proper range Pass/Fail
8. Exercise test
- Start manual exercise test, e.g. initial load of 50 Watts / load change of 25 Watts.
- Check load level and load indication.
Expected result: Load changes as expected Pass/Fail
9. Communication (if applicable)
- If ergometer is remote-controlled from EKG unit or PC-ECG system, check connection cable.
- Check load control vial EKG unit or PC ECG system.
- Check remote start if applicable.
Expected result: Ergometer can be controlled by external unit Pass/Fail
A: Final Checkout Procedure / Functional Test