2018112-003 Rev J eBike, eBike L, eBike EL - 53 -
D: Service Menu
* Push softkey SET to confirm the value. A reset of the MRD
board is performed and the current offset voltage is stored. This
may take a few seconds.
* When done, the screen shown at right appears. You are now
asked to attach the 8 kg weight.
The display shows the WSG voltage for that 8 kg weight. It must
be greater than 800 mV.
With the arrow keys, the displayed weight can be adjusted to
8 kg (range from 7 to 9 kg in steps of 0.1 kg). When you press
an arrow key, the weight is displayed in reverse video.
* This adjustment is accepted when pressing the SET key. The
corresponding value will be stored. Wait for some seconds.
* Confirming the adjustment
- After successful calibration a short beep sounds and the
message „Calibration successful“ is displayed.
- If calibration fails, a long beep sounds and the message
„Calibration failed“ is displayed.
* Press the Menu key to exit the calibration mode and return to
the Service menu.
* To exit the Service menu, you have to switch the ergometer OFF
and ON again.
Remove the weight before switching the ergometer on
again.
Note: If you turn the ergometer on again and the weight is still
attached, the error code E:02 will appear. The unit is locked
and will not enter the normal operating mode.
Load Calibration
Wait 3 . . 6 sec.
50
0
Load Calibration
Load 8 kg weight
DMS-Voltage: 1012 mV
Load: 8 kg
Load Calibration
Wait 3 . . 6 sec.
50
0
Load Calibration
DMS-Voltage 1019 mV
Load 8.0 kg
50
0
Calibration successful
Load Calibration
DMS-Voltage 63 mV
Load 0.0 kg
50
0
Calibration Failed