Guidance and Manufacturer’s Declaration – Emissions
The enFlow 100 with enFlow 121 is intended for use in the electromagnetic environment specied below. The customer or user of the enFlow 100 with enFlow 121 should
ensure that it is used in such an environment.
Immunity Test EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile communications equipment should be separated from
the enFlow 100 with enFlow 121 by no less than the distances calculated/
listed below where P is the max power in watts and D is the recommended
separation distance in meters:
D=(3.5/3V/m)(Sqrt P)
D=(3.5/3V/m)(Sqrt P)
80 to 800 MHz
D=(7/3V/m)(Sqrt P)
800 MHz to 2.5 GHz
Field strengths from xed transmitters, as determined by an electromagnetic
site survey, should be less than the compliance levels (3Vrms and 3V/m).
Interference may occur in the vicinity of equipment containing a transmitter.
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the enFlow
The enFlow 100 with enFlow 121 is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the enFlow 100
with enFlow 121 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the enFlow
100 with enFlow 121 as recommended below, according to the maximum output power of the communications equipment.
Max Output Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
Separation (m)
80 to 800MHz D=(3.5/E1)
(Sqrt P)
Separation (m) 800MHz to 2.5GHz D=(7/E1)(Sqrt P)
0.01 .1166 0.12 0.23
0.1 .3689 .3689 .7378
1 1.1666 1.1666 2.3333
10 3.6893 3.6893 7.3786
100 11.6666 11.6666 23.3333