- 26 of 40 -
4400-0024-EN Rev A
Appendix C: Warming system response by temperature
Fluid Temp Heater Temperature LED on Warmer Display on Controller Audible Alarm
30 °C Active Blue Flashing 30 °C Blue Flashing No
31 °C Active Blue Flashing 31 °C Blue Flashing No
32 °C Active Blue Flashing 32 °C Blue Flashing No
33 °C Active Blue 33 °C Blue No
34 °C Active Blue 34 °C Blue No
35 °C Active Green 35 °C Green No
36 °C Active Green 36 °C Green No
37 °C Active Green 37 °C Green No
38 °C Active Green 38 °C Green No
39 °C Active Green 39 °C Green No
40 °C Active Green 40 °C Green No
41 °C O Green 41 °C Green No
42 °C O Green 42 °C Green No
43 °C O Yellow 43 °C Yellow No
44 °C O Yellow 44 °C Yellow No
45 °C O Red Flashing High Priority Alarm after 20 seconds 45 °C after 20 seconds Red Flashing “Over Temp” message After 20 seconds
46 °C O Red Flashing High Priority Alarm after 16 seconds 46 °C after 16 seconds Red Flashing “Over Temp” message After 16 seconds
47 °C O Red Flashing High Priority Alarm after 12 seconds 47 °C after 12 seconds Red Flashing “Over Temp” message After 12 seconds
48 °C O Red Flashing High Priority Alarm after 8 seconds 48 °C after 8 seconds Red Flashing “Over Temp” message After 8 seconds
49 °C O Red Flashing High Priority Alarm after 4 seconds 49 °C after 4 seconds Red Flashing “Over Temp” message After 4 seconds
50 °C O Red Flashing High Priority Alarm (immediately) Red Flashing Over Temp message (immediately) Immediately