4400-0024-EN Rev A
5. Insert the enCheck unit into the Warmer. The end of the unit is keyed similar to the Cartridge so it will only t in the correct orientation. Close the covers.
POWER
TEMP
OFF
ON
6. Conrm the enCheck is set to the normal mode.
7. Move the MAINS power switch on the back of the Controller to the ON position. Wait for the thermometer to stabilize, ≈ 30 to 60 sec. assuming all
equipment is close to 20 °C.
8. The temperature on the thermometer should be 40 °C ± 2 °C.
Note : If the temperature in step 8 above is not 40 °C ± 2 °C, take the Warmer unit out of service. Call Customer Service for an RGA. Please reference the
RGA section in “Servicing the enFlow IV uid/blood warming system components” for further information regarding returns.