4400-0024-EN Rev A
enFlow
Warm er 
T hermometer
Temperature
probe
inserted  into
extension set.
Waste water
container
IV Line set
Extension set
enFlow
Controller 
Power
source
Pump
100
mL/min
Temperature
probe
inserted  into
extension set.
Extension set
.5 L
Water source
50 °C
3.  Change the source uid temperature to 50 °C ± 2 °C. Take extra care while working with hot uids. 
4.  Power on the enFlow system. 
5.  Allow at least 20 seconds for the power-on self-test to complete and the temperature display to ready a stable temperature. Conrm all LEDs are 
illuminated green. 
6.  Release the clamp and establish a uid ow of 100 ± 20 mL/min. 
7.  With the thermometer, conrm the temperature of the uid at both the input and output end of cartridge. 
8.  The Over-Temperature alarm sounds within approximately 20 seconds or less of the output uid temperature reaching that of the input uid level. 
A.  Conrm the Controller produces an audible beep and displays a red Over-Temp message on the display (Controller only). 
B.  Conrm the temperature LED on the Warmer ashes red indicating that the uid has gone over-temperature.