4400-0024-EN Rev A
enFlow
Warm er
T hermometer
Temperature
probe
inserted into
extension set.
Waste water
container
IV Line set
Extension set
enFlow
Controller
Power
source
Pump
100
mL/min
Temperature
probe
inserted into
extension set.
Extension set
.5 L
Water source
50 °C
3. Change the source uid temperature to 50 °C ± 2 °C. Take extra care while working with hot uids.
4. Power on the enFlow system.
5. Allow at least 20 seconds for the power-on self-test to complete and the temperature display to ready a stable temperature. Conrm all LEDs are
illuminated green.
6. Release the clamp and establish a uid ow of 100 ± 20 mL/min.
7. With the thermometer, conrm the temperature of the uid at both the input and output end of cartridge.
8. The Over-Temperature alarm sounds within approximately 20 seconds or less of the output uid temperature reaching that of the input uid level.
A. Conrm the Controller produces an audible beep and displays a red Over-Temp message on the display (Controller only).
B. Conrm the temperature LED on the Warmer ashes red indicating that the uid has gone over-temperature.