• Avoid puncturing the heating surface. If damaged, remove the Warmer from service and replace immediately.
• This equipment is not intended for use in an oxygen rich environment (dened as >22% O
2
).
• No modication of this equipment is allowed.
• Do not position the device in a way that makes it dicult to disconnect the device.
• Due to highly stable components, microprocessor control, and built-in self-tests, an annual performance check is sucient.
• The steps listed in the enFlow IV uid/blood warming system operational checklist (Appendix F) should be performed at least once a year, or as required by
your accrediting body.
• The enCheck is to be used only with the enFlow Warmer and Controller.
• Do not touch the enCheck contact plate surface during or immediately after use since it may be very hot.
• The warmer and controller may be a potential biohazard during or after use. Handle and dispose of in accordance with acceptable medical practice and
applicable regulations at the end of their life.