ActionFault
The drop sensor photocell is dirty. Clean the photocell
with a damp cloth.
Check that the capillary end projection is not too long
(~2 mm).
Check that the flow rate is not too high (a continuous
flow).
DropSync is not func-
tioning
Check that the safety bar is folded up.No fractions are collect-
ed
Check that the DropSync unit is close enough over the
tubes.
Check that the rack is correctly fitted to the rack holder.
Check that the capillary end is cut cleanly and straight.
Check that the correct rack type is selected.
Liquid misses the tubes
50 Frac-920 and Frac-950 Operating Instructions 28-9622-09 AC