GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 58 Section 7-6 - Service Desktop (R4.2.5x and later)
To configure automatic data transfers:
1.) Navigate to ADMIN > System Admin.
2.) On System Admin, configure these settings:
- Transfer Machine Logs to GE – Default is checked
- Transfer Operator Information to GE
- Transfer Images (captured without patient information) to GE – Default is checked.
3.) Click Save.
4.) Navigate to Utility (second page) > Service > Utilities > Data Transfer.
5.) On Data Transfer, select Scheduler, and then select the days to perform the data transfer.
6.) Click Save Settings.
7.) To manually perform a data transfer, click Send All.