GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-5 - Preparing for setup 3 - 11
Section 3-5
Preparing for setup
3-5-1 Verify Customer Order
Compare all items listed on the packing slip (delivery note) with those received.
Report any items that are missing, back ordered, or damaged.
Figure 3-5 envelope at front panel of the crate
3-5-2 Physical inspection
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to 3-3-2-3 "Damage in transportation" on page 3-6 in the
beginning of this manual.
3-5-3 EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
Refer to Section 2-2-3 "EMI Limitations" on page 2-4.
Check whether delivery is complete (according to packing list) and check visual
damage!
The envelope with delivery address, packing list and invoice is located on the
front panel of the crate.

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals