EasyManuals Logo

GE LOGIQ E8 User Manual

GE LOGIQ E8
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #353 background imageLoading...
Page #353 background image
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 83
7-7-20 When the touch panel does not response
From R4 system, the touch panel changes to electrostatic type. If the keys does not respond, clean the
touch panel first. If the gel or somehting remains on the touch panel, it does not reposnd.
7-7-21 Limitation message during AFI Analysis
The limitaion message “Over Depth too high for analysis xx.xx cm, EXIT" displays when you measure
AFI deeper than 20 cm. Measure whithin 20cm.

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals