GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-5
8-2-1-2 Removal procedure (cont’d)
6.) Unscrew 2 screws to remove the OSD key assy and disconnect the connect.
7.) Separate the 17" LCD tact knob from the OSD key assy.
Figure 8-6 Separating the OSD key assy
Figure 8-7 Separating the 17" LCD tact knob