EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V3

GE LOGIQ V3 User Manual

GE LOGIQ V3
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
General Procedures and Functional Checks
4-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
4-2-7 Backup and Restore Database, Preset Configurations and
Images
NOTE: Always save presets before any software reload. This ensures
the presets loaded after the software reload are as up–to–date
as possible.
All user presets except changes to Summary, Anatomy, and
Biometry pages, can be saved on an DVD-R disk (or USB
memory device) for reloading on the system.
NOTE: Presets should NOT be saved on the same USB memory device
(or DVD-R disk) as images. The Archive Menu lists the images
but does NOT list the presets stored on a USB memory device
(or DVD-R disk).

Table of Contents

Other manuals for GE LOGIQ V3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V3 and is the answer not in the manual?

GE LOGIQ V3 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V3
CategoryMedical Equipment
LanguageEnglish

Related product manuals