CARESCAPEV100
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohard-to-cleanareas,likegroovesandcrevices.
Removeanysoilassoonaspossibleafterusebywipingthedevice.
Anycontactofcleaningsolutionswithmetalpartsmaycausecorrosion.
Donotdamageorbendconnectorpinswhencleaningordrying.
Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
6.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedcloth.
7.Visuallyensurethatthedeviceisclean.Ifvisiblesoilremains,repeatcleaning
procedureuntilthedeviceisthoroughlyclean.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
9.Reconnectthedevicetothepowersupply.
10.Reconnectallcables.
11.Turnonthepowertothedevice.
Othermanufacturercleaninginstructions
Fordetailsaboutcleaninganddisinfectioninformationfordevices,device
components,supplies,andaccessoriesmadebymanufacturersotherthanGE,see
theapplicableinstructionsforuseprovidedbythemanufacturer.
Donotreusesingle-usedisposableaccessories.
Disinfection
Disinfectioncleaningprecautions
Disinfectionnotices
NOTICE
SPOTTING,STAINING,CORROSION.Metalcomponents
includingfastenersmaycorrodeorbecomestainedor
spottedinappearancefromcontinuousexposuretochlorine
compoundslikebleach.Theseeffectsarenotassociatedwith
productsafetyornormalfunction.
Disinfectionpointstonote
Alwayscleanbeforedisinfecting.
Alwaysdilutecleaninganddisinfectantagentsaccordingtotheirmanufacturer’s
instructions.Alwaysconsideryourhospitalguidelinesaswell.
Useonlythepermittedsubstances.
Yourhospitalguidelinespermitting,allcleaninganddisinfectionactivitiescan
becarriedoutatthebedside.
Beforedryingasurfaceafterwipingit,waittheminimumtimerequiredaccording
tothesubstancemanufacturer’sinstructions.
Visuallyensurethatnosubstanceresidueremainsonthedevice.