PatientMonitorB105/B125
WARNING
Neveruseconductivesolutions,oxidizingcompounds,wax,or
waxcompoundstocleandevices,cablesorleadwires.
Cleaningcautions
CAUTION
Donotapplypressurizedairorgastoanyoutletortubing
connectedtothemonitor.Pressuremaydestroysensitive
elements.
Permitteddetergents
Thefollowingpermitteddetergentscanbeusedtocleanthedeviceandother
non-appliedpartsunlessthereareseparatepart-specicinstructions.Non-applied
partsrefertothosepartsofthesystemthatarenotindirectcontactwiththepatient.
Water
EnzymaticorneutralpHdetergents
Cleaningpointstonote
WarrantydoesnotcoveranydamagescausedbyusingotherthanGEapproved
substancesandmethods.
GEmakesnoclaimsastotheefcacyofthelistedsubstancesormethodsto
controlinfection.
Alwaysdilutecleaninganddisinfectantagentsaccordingtotheirmanufacturer’s
instructions.Alwaysconsideryourhospitalguidelinesaswell.
Yourhospitalguidelinespermitting,allcleaninganddisinfectionactivitiescan
becarriedoutatthebedside.
Donotletliquidpoolaroundconnectionpins.Ifthisshouldhappen,blotdrywitha
soft,lint-freecloth.
Donotuseexcessivedryingtechniques,suchasoven,forcedheat,orsundrying.
Donotspraycleanerdirectlyonthedisplayscreen.
Neverconnectanydeviceorappliedparttoapatientuntilitisthoroughlydry.
Ifyoudiscoveranysignsofdeteriorationordamageinthedevice,discontinueits
use.
Cleaninginstructions
Lockingthescreen
Youcansetthetouchscreenfeatureoffwhenyouneedtocleanthescreen.
1.Select
> Lock.
2.Toenablethetouchscreen,pressanykeypadkey,orusetheTrimKnob.
NOTE
PresstheOn/Offkeytoturnoffthemonitor.