CARESCAPEB650Monitor
WARNING
Donotautoclaveanypartofthesystemwithsteam(including
cablesorleadwires)orsterilizewithethyleneoxideor
radiation.
WARNING
Donotpourorsprayanyliquidsothatitmayseepinto
connectionsoropenings.
WARNING
Neveruseconductivesolutions,oxidizingcompounds,wax,or
waxcompoundstocleandevices,cablesorleadwires.
Cleaningcautions
CAUTION
Donotapplypressurizedairorgastoanyoutletortubing
connectedtothemonitor.Pressuremaydestroysensitive
elements.
Permitteddetergents
Water
Ethanol70to96%byvolume
Isopropylalcoholmax.60%byweight
EnzymaticorneutralpHdetergents
Cleaningpointstonote
WarrantydoesnotcoveranydamagescausedbyusingotherthanGEapproved
substancesandmethods.
GEmakesnoclaimsastotheefcacyofthelistedsubstancesormethodsto
controlinfection.
Yourhospitalguidelinespermitting,allcleaninganddisinfectionactivitiescan
becarriedoutatthebedside.
Donotletliquidpoolaroundconnectionpins.Ifthisshouldhappen,blotdrywitha
soft,lint-freecloth.
Donotuseexcessivedryingtechniques,suchasoven,forcedheat,orsundrying.
Donotspraycleanerdirectlyonthedisplayscreen.
Neverconnectanydeviceorappliedparttoapatientuntilitisthoroughlydry.
Ifyoudiscoveranysignsofdeteriorationordamageinthedevice,discontinueits
use.
Settingthetouchscreenoffforcleaning
Youcansetthetouchscreenfeatureofffor30secondsatatimewhenyouneedto
cleanthescreen.Thecountdowntimerappearsonthescreen.
1.SelectMonitorSetup>MainSetup.
2.SelectTouchscreenOff.
3.Youcanenablethetouchscreenimmediatelybypressinganymonitorhardkeyor
theTrimKnob,orbyselectingCancelintheTouchscreenOffdisplay.