ROOM REQUIREMENTS
GE Healthcare SENOGRAPHE DMR+
REV 1 pim 5128706–1–100
2–5
3–5 Line supply cable
The line supply cable comprises of two supply wires and a ground cable (i.e. 3 x AWG 10 (5.32
mm
2
)) with the following actual/usable lengths:
1+0 0)<+2 7 $ ,D-9
%/E0) 0)<+2 7 F $ ,-
%&
  +
3–5–1 Obtaining a line supply cable
D For customers in all countries except China, the optional line supply cable (S30331BC)
must be ordered from the Price Book so that it is supplied with the Senographe System.
D For customers in China, the optional line supply cable from GEMS must not be ordered.
Instead Chinese customers must locally order an equivalent CCC-certified supply cable
from their local electrical supplier.
3–6 Main circuit breaker
0 ONLY for GEMSE:
the Generic EDM is associated with an external modem (Motorola
3265) mounted on the EDM cabinet.
For GEMS A and GEMS AM the Generic EDM incorporates a modem board inside the
EDM cabinet. 2) $ 3*34+ E*)@)* / /4((0); E= +2) 34/+1$)* ; $4/+ E)
/G);  331*;3) 8+2 0130 *)<40+1/
Circuit breakers sizes for European market:
H*1$ 6 4( +1 " > 3*34+ E*)@)*>  7  $<)+3 7  '9
H*1$  4( +1 " > 3*34+ E*)@)*>  7 
$<)+3 7  '
Circuit breakers size and supply conductors for US market: refer to Section 517–71(a)
and Section 517–73(a) (Item 1, 2) of the NEC–1993 (see below).