ROOM REQUIREMENTS
GE Healthcare SENOGRAPHE DMR+
REV 1 pim 5128706–1–100
2–8
4–3 Electromagnetic Immunity
The Senographe is suitable for use in the specified electromagnetic environment. The purchaser or
user of the Senographe should ensure that it is used in an electromagnetic environment as described
below:
TABLE 2–2
ELECTROMAGNETIC IMMUNITY –PART 1
Immunity Test IEC
60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors are wood, concrete, or ceramic tiles, or floors are cov-
ered with synthetic material and the relative humidity is at
least 30 percent.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power supply
lines
1 kV for
input/output
lines
2 kV for
power supply
lines
1 kV for
input/output
lines
Mains power quality is that of a typical commercial and/or
hospital environment
Surge
IEC 61000-4-5
1 kV differen-
tial mode
2 kV com-
mon
mode
1 kV differen-
tial mode
2 kV com-
mon
mode
Mains power quality is that of a typical commercial and/or
hospital environment.
Voltage dips,
short interrup-
tions and volt-
age variations
on power sup-
ply input lines
IEC
61000-4-11
0% U
n
for 5
sec
0% U
n
for 5
sec
Mains power quality is that of a typical commercial and/or
hospital environment. If the user of the Senographe requires
continued operation during mains power interruptions, it is
recommended that the Senographe be powered from an
uninterruptible power supply or a battery.
Power fre-
quency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
1A/m
Power frequency magnetic fields are at levels characteristic
of a typical location in a typical commercial and/or hospital
environment.
At this disturbance level the monitor image may present some
slight flicker. If this occurs, the monitor may be removed from
the vicinity of the low frequency magnetic field source to im-
prove image quality.
At this disturbance level the system is fully operational.
Note: These are guidelines. Actual conditions may vary.