ROOM REQUIREMENTS
GE Healthcare SENOGRAPHE DMR+
REV 1 pim 5128706–1–100
2–10
4–4 Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC 60601-1-2
TABLE 2–4
RECOMMENDED SEPARATION DISTANCES
Frequency of
Transmitter
Equation
150KHz to 26
MHz
d = (3.5/V
2
).P
26 MHz to 80 MHz
d = (3.5/V
2
) .P
80 MHz to 800
MHz
d = (3.5/E
2
).P
800 MHz to 2.5
GHz
d = (7/E
4
).P
Rated Power
of Transmitter
(watts)
DISTANCE
(meters)
DISTANCE
(meters)
DISTANCE
(meters)
DISTANCE
(meters)
10 mW 1.2 1.2 1.2 2.3
100 mW 3.8 3.8 3.8 7.3
1 12 12 12 23(*)
10 38 38 38 73
100 120 120 120 230
D For transmitters rated at a power not listed above, the recommended separation distance
(d, in meters) can be estimated using the equation in the corresponding column, where
P is the power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
D (*) For example, a 1 W mobile phone (800MHz to 2.5GHz carrier frequency) should be
no closer than 23 meters from the Senographe to avoid image interference risks.
D Using the recommended distance as determined from Table 4, between 150KHz &
2.5GHz, some slight disturbance might be observed at image level. The disturbance to
the image cannot be confused with a medical pathology, and the equipment retains its
medical utility.
Note: These are guidelines. Actual conditions may vary.
4–5 Use Limitation
D External components:
The use of accessories, transducers, and cables other than those specified may result in
degraded Electromagnetic compatibility of the Senographe.
4–6 Installation Requirements and Environmental Control
D In order to minimize interference risks, the following requirements apply.
4–6–1 Cable shielding & grounding
D All interconnecting cables to peripheral devices must be shielded and properly grounded.
Use of cables not properly shielded and grounded may result in the equipment causing
radio frequency interference.
4–6–2 Separated power supply distribution panel & line
D This product complies with the radiated emission limits of the CISPR11 Group1 Class
A standard.

Other manuals for GE SENOGRAPHE DMR+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE SENOGRAPHE DMR+ and is the answer not in the manual?

GE SENOGRAPHE DMR+ Specifications

General IconGeneral
BrandGE
ModelSENOGRAPHE DMR+
CategoryMedical Equipment
LanguageEnglish

Related product manuals