ROOM REQUIREMENTS
GE Healthcare SENOGRAPHE DMR+
REV 1 pim 5128706–1–100
2–12
SECTION 5
INSITE CONNECTION
The Senographe DMR will use the Generic EDM (Equipment Diagnostic Monitor) when it is
available (1st quarter 1996). The Generic EDM is supplied as a Service option, in a separate
cabinet. The Generic EDM can be installed on a shelf or anchored to the wall using an
installation kit. The Generic EDM will be preferably installed close to the DMR generator
cabinet but as the Generic EDM can monitor several systems at the same time, it can also be
placed in another room. See Generic EDM manual for additional details.
The Generic EDM cabinet has the size of a mini–tower PC:
Height x Lenght x Width = 350 x 430 x 170 in mm = 14 x 17 x 7 in inches.
A dedicated phone line with a local socket used only for the connection to a modem will
preferably be located close to the EDM cabinet. The Generic EDM is supplied with 220V AC
exclusively. This supply must be as permanent as possible (e.g. not switchable with the main
circuit braker of the DMR).
0 ONLY for GEMSE:
the Generic EDM is associated with an external modem (Motorola
3265) mounted on the EDM cabinet.
For GEMS A and GEMS AM the Generic EDM incorporates a modem board inside the
EDM cabinet.

Other manuals for GE SENOGRAPHE DMR+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE SENOGRAPHE DMR+ and is the answer not in the manual?

GE SENOGRAPHE DMR+ Specifications

General IconGeneral
BrandGE
ModelSENOGRAPHE DMR+
CategoryMedical Equipment
LanguageEnglish

Related product manuals