Chapter 15 Page no. 164 15-maint.fm
GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Maintenance
3-7 Evaluation
• For the flexible compression paddle, which is not intended to be flat and parallel to the breast support
table during compression, the deflection must be within the following values:
- Minimum: 12 mm.
- Maximum: 14 mm.
3-8 Corrective action
• If the deflection exceeds the specified limits, the paddle should be replaced.
3-9 References
1. I. Fife, "The physical dimensions of the compressed breast," British Jnl. of Radiology, 64, 73,
1991.
2. Handbook of Glandular Tissue Doses in Mammography, HHS Pub. FDA 85-8239, U. S. Depart-
ment of Health and Human Services, Food and Drug Administration, Rockville, MD, March, 1985.
3. Keith Nelson, Standard Imaging, Middleton, WI, private communication.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).