GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
rev_ENG1.fm Page no. 205
Revision History
This page is intentionally left in English in the translated versions.
REV DATE REASON FOR CHANGE
1 January 25, 2005 Release Nephtys M5B level systems. Based on document 5131764-2-
1xx rev 1, with the following changes:
- Ch 16 Error Messages, Table 3 :
- Neph mess 449 "1.5 Mag stand present. Remove face shield!" modi-
fied
- Neph mess 448 "1.8 Mag stand present. Remove face shield!" modi-
fied
- Neph mess 438, Neph mess 434 "2D Large Localization Paddle.
Maximum recommended force for this paddle is 20daN." modified
- Neph mess 442 "Round Spot Magnification paddle" modified
- Neph mess 430 "19x23 paddle" modified
- Neph mess 429 "Flexible 19x23 paddle" modified
- Neph mess 435 "2D Spot Localization Paddle. Maximum recom-
mended force for this paddle is 20daN." modified
- Neph mess 443 "Square 7x7 Magnification paddle" modified
- Neph mess 440 "19x23 Magnification paddle" modified
- Neph mess 1261 "H32 Tube head control failure" added
- Neph mess 1262 "H34 Tube head communication failure" added
- Neph mess 1265 "W32 Take care, floor anti-collision is disabled"
added
- Neph mess 466 "S59 Gantry device config. not allowed" Explanation
and Corrective Action modified
- Neph mess 465 "S60 X-ray beam adjustment error" replaced by "S60
X-ray beam quality error", Explanation and Corrective Action modified
- Neph mess 464 "S61 X-ray beam not ready" Explanation and Correc-
tive Action modified
- Neph mess 436 "Unknown paddle type" Corrective Action modified
- Neph mess 383 "X13 X-ray beam adjustment error" Explanation and
Corrective Action modified
- Neph mess 1231 "W29 Paddle detection error" added
- Neph mess 252 "H15 Stereotactic Pos. comm. Failure" added
- Neph mess 1255 "A26 Something is preventing decompression
movement" added
- Neph mess 1256 "W31 No defined paddle detected" added
- Ch 15 Maintenance, Section 2-1 Planned maintenance performed by
the Field Service Engineer, added planned maintenance calendar. Also,
in first sentence Section 2-1: replaced should by shall.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals