Chapter 16 Page no. 166 16-errors.fm
GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
Table 1 Error Messages displayed on the X-ray Console
Message on
X-ray Console
Explanation Corrective Action
E01 Generator failure Problem reported in HV section
of generator.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E02 Focus bias failure Problem reported in tube focus
bias circuits.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E03 Rotor failure Problem reported in anode
starter.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E04 Arm fatal failure Generator detects error in dialog
with arm.
Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
E05 Failure during
exposure
Current exposure aborted due to
failure in equipment.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
inform GEMS Field Service.
E06 Software error (gen
CPU)
An error occurred in the Genera-
tor CPU software.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
inform GEMS Field Service.
E07 Please TURN OFF
the console !
A generator emergency failure
occurred but a second failure
prevented the system from shut-
ting itself down.
1) Note ALL messages.
2) Switch OFF the system from the X-ray console
to quickly switch off the generator.
3) Contact GEMS Field Service.
E08 Power supply fail-
ure
Supply of generator power cir-
cuits is not authorized.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E10 Heating failure
track 1
Molybdenum focal track is unser-
viceable.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL
messages and contact GEMS Field Service.
You can still work with rhodium focal track.
E11 Heating failure
track 2
Rhodium focal track is unservice-
able.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
You can still work with molybdenum focal track.
E12 Checksum error
(gen)
Calibration parameters have
been modified.
Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
E13 Gen./Arm commu-
nication fail
Generator cannot communicate
with arm.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact
GEMS Field Service.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals