GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm Page no. 173 Chapter 16
S09 No compression
detected
Breast is not compressed. Breast compression is mandatory for AOP.
S10 Generator cooling
mn
Generator requires xxxx minutes
to cool.
Wait for cooling period (or select lower mAs values
in Manual Mode).
S100 PLEASE WAIT ... Generator requires xxxx minutes
to cool.
Wait for cooling period (or select lower mAs values
in Manual Mode).
S11 Tube cooling mn Tube requires xxxxx minutes to
cool.
Wait for cooling to terminate (or select lower mAs
values in Manual Mode).
S12 Configuration not
allowed
Incompatible settings of magnifi-
cation, grid, and focal spot have
been selected.
Select compatible settings.
S13 Backup mAs < 50
exposure inhibited>
Maximum mAs < 50.
Exposure inhibited.
Wait for cooling to terminate (or select lower kV
values).
S14 Maximum mAs
exceeded
Maximum mAs set at installation
are exceeded. This message
appears only if a value below the
maximum possible mAs value
was set at installation.
Reduce the mAs value.
S20 Acq. Abort, image
can be lost
Exposure has been fully per-
formed but image may be lost
because an error occurred in the
image chain detection or pro-
cessing.
1. Press the Exposure Interrupt button.
2. Check image quality if image is displayed.
3. Start exam again.
4. If the problem persists, power OFF the whole
system and wait for the end of the shutdown
sequence.
5. Restart the system.
6. If the problem occurs again, note ALL messages
and inform GEMS Field Service.
S21 Waiting for AWS
boot
Communication with AWS sys-
tem is not yet established.
1. Wait until AWS boot and login is finished.
2. If message persists 10 min after the AWS login,
power OFF the whole system and wait for the end
of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
S22 Waiting for IDC
boot
Communication with Image
Detection Control system is not
yet established.
1. Wait until the Image Detection Control boot is
finished.
2. If message persists for more than 10 min, power
OFF the whole system and wait for the end of the
shutdown sequence.
3. Restart the system.
4. If the problem occurs again, note ALL messages
and inform GEMS Field Service.
S23 Laterality is not
selected
You have not selected the lateral-
ity of the breast to perform acqui-
sition. Exposure is inhibited.
Select laterality on the X-ray Console. If you are
imaging a test object, the laterality chosen is not
important.
Message on
X-ray Console
Explanation Corrective Action
FOR TRAINING PURPOSES ONLY!